S. No |
Description |
Specification |
1 |
Scanning Electron Microscopy (SEM) with XMA Analyzer |
Field Emission SEM: Accelerating voltage 30 kV, Magnification up to 1,000,000X EDX system with software library (liquid N2 free) Low vacuum mode pressure range 10 Pa to 300 Pa A stage, which holds a sample size of at least 50mm x 50 mm; Resolution range=0.8nm at 15kV. |
2 |
X-ray Diffractometer (XRD) |
Theta/theta (ie fixed sample) with 9 position sample changer, position sensitive detector (LynxEye) and standard detector (SC) with auto-absorber & graphite 2rd beam monochromator and sample rotation. Divergent beam, Bragg Brentano parafocusing geometry, reflection mode and most slit motorized. (min step 0.0001° for detector & 0.0001° for tube (omega), best instrumental resolution ~0.08° in 2theta & ~0.04° in omega). For Thin film/powder samples XRD analysis software facility with library |
3 |
Fuel cell |
APPROXIMATE POWER RANGE: 1 – 50 W GAS FLOW: 100:1 turndown MFC’s Standard anode flow range: 0.015 – 1.5 nlpm Standard cathode flow range : 0.025 – 2.5 nlpm GAS HUMIDIFICATION TECHNOLOGY: - Dew Point control: Up to 95°C (203°F) optional high temperature operation Gas Temperature: Up to 110°C (203°F) optional high temperature operation Dry gas by-pass: optional; with automatic control ELECTRICAL END PLATE HEATER CONTROL: include; with software PID BACK PRESSURE CONTROL: Up to 3barg (45psig); higher pressures available DIMENSIONS (L x W x H) (approximate): 890x710x1855 mm (32x28x73 in) Effective area of cell : 5cm² |
4 |
Au Sputter Coater |
Mini plasma sputtering coater with vacuum pump and gold target |
5 |
Spin Coater |
Programmable Desktop Spin Coater Variable Speed : 0 – 8000rpm Natural polypropylene housing with interlocking Electronic programmable controller 13100 – 12000 rpm with resolution of <0.5 rpm Vacuum pump |
6 |
Spectrophotometer UV/Vis/NIR |
Range (175 nm to 3300 nm) PMT, InGaAs and PbS 3-detector module High sensitivity resolution 0.2 nm Snap-in integrating spheres to 150 mm Precision LCR Meter 20 Hz to 2 MHz Vector Network Analyzer (3 GHz to 20 GHz) |
7 |
Metallographic Sample Preparation (Sample cutter, Mounting Press and Automatic Polisher) |
Sample Cutter: Grinding wheel cutter Mounting Press: Bakelite mounting samples Rough Grinding machine (Wheel Dia 250 mm): Each Machine with two Wheels along with holding rings, Variable Speed Control, Spare wheels with all supplies (200 each Non-stick Emery Papers 120#, 200#, 400#, 800#,1200# of 250 diameter) Polishing Machine Wheel Dia 250 mm: Each Machine with two Polishing Wheels along with holding rings, Variable Speed Control, Spare wheels with all supplies (200 each both nap less and with nap polishing cloth, 5 diamond paste tubes (6 micron, 1 micron), alumina polishing powders 5 Kg each (10micron, 1micron, 0.05 micron) |
8 |
Universal Testing Machine 30 kN |
Electromechanical, Capacity approx. 30KN (with tensile, bending & compression and fiber test options). Loads cells: 5KN, 20KN and 30KN Grips: For Fiber, sheets, rectangular flat, and rod samples Options: Optical extensometer to measure the gauge length, specimen making accessories and mold according to the standards for plastics to cut dumb-bell test pieces. |
9 |
Universal Hardness Tester |
Capable of the following hardness tests: Vicker, Rockwell and Brinell |
10 |
Differential Scanning Calorimetry (DSC) |
Performance Temperature Range Ambient to 725°C With Cooling Accessories -180 to 725°C Temperature Accuracy +/- 0.l°C Temperature Precision +/- 0.0l°C Calorimetric Reproducibility (indium metal) +/- 0.05% Calorimetric Precision (indium metal) +/- 0.05% Dynamic Measurement Range >+/- 500 m-W Baseline Curvature (Tzero; -50 to 300•C) 10 µW Baseline Reproducibility with Tzero +/- 10 µW Sensitivity 0.2 µW Indium Height/Width 60 mW /°C |
11 |
Dynamic Mechanical Analyzer (DMA) |
Maximum Force 8 N Minimum Force 0.0001 N Force Resolution 0.00001 N Strain Resolution 1 nanometer Modulus Range 103 to 3x 1012 Pa Modulus Precision ± 1% Tand Sensitivity 0.0001 Tand Resolution 0.00001 Frequency Range 0.01 to 200 Hz Dynamic Sample Deformation Range ± 0.5 to 10000 µm Temperature Range -150 to 600°C Heating Rate 0.1 to 20°C/min Cooling Rate 0.1 to 10°C/min Isothermal Stability ±0.1°C Time/Temperature Superposition YES RH Control Optional |
12 |
Differential Thermal Analysis System (DTA) |
System Design Horizontal Balance & Furnace Balance Design Dual Beam (growth Compensated) Sample Capacity 200 mg (350 mg including sample holder) Balance Sensitivity 0.1 µg Furnace Type Bifilar Wound Temperature Range Ambient to 1500C Heating Rate - Ambient to 1000C 0.1 to 100C/min Heating Rate - Ambient to 1500C 0.1 to 25C/min Furnace Cooling Forced Air (1500 to 50C in < 30 min, 1000C in 50C in < 20 min) Thermocouples Platinum/Platinum-Rhodium (Type R) Temperature Calibration Curie point or Metal Standards [1 to 5 Points) DTA Sensitivity 0.001C Calorimetric Accuracy/Precision ± 2% [based on metal standards) Mass Flow Controller with Automatic Gas Switching Included Vacuum to 7 Pa (0.05 torr) Reactive Gas Capability Included - separate gas tube Dual Sample TGA Included Auto-Stepwise TGA Included Sample Pans Platinum: 40 µl, 110 µl Alumina: 40 µl, 90 µl |
13 |
Thermogravimetric Analysis (TGA) with Evolved Gas Analysis Furnace (EGA) |
Temperature Compensated Thermobalance Included Maximum Sample Weight l g Weighing Precision +/- 0.01% Sensitivity 0.1 µg Baseline Dynamic Drift* < 50 µg Furnace Heating Resistance Wound EGA Furnace Optional Temperature Range Ambient to 1000°C Isothermal Temp Accuracy +/- 1°C Isothermal Temp Precision +/- 0.1°C, Controlled Heating Rate 0.1 to 100°C/min Furnace Cooling (forced air/N2) 1000 to 50°C < 12 min Temperature Calibration Curie Point Auto-loader Included Auto Stepwise TGA Included TGA/MS Operation Optional TGA/FTIR Operation Optional PlatinumTM Software Included Sample Pans Platinum50, 100 µl Ceramic 100, 250, 500 µl
*From 50 to 1000C at 20C/min using empty platinum pons, no baseline/blank subtraction. |
14 |
FTIR (Fourier Transform Infra-red Spectroscopy) |
Interferometer: 30 degree incident angle Michelson interferometer with dynamic alignment system. Spectral Range: 350 to 7800 /cm expandable to 12500/cm. Resolution: not more than 0.25/cm Detector: High sensitive detector DLATGS
S/N ratio: 60,000 or higher Having future up-gradation to use with IR microscope.
Sampling accessories: 1. DRS diffused reflectance attachment 2. ATR attenuated total reflectance attachment 3. Specular Reflectance attachment for analysis of thin film on metal with um order of thickness. 4. Fixed and demountable liquid cell. |
15 |
Control Unit and Ultra Micro balance |
AD6000-Control Unit and Ultra microbalance module, 0.1 microgram sensitivity |
16 |
Impact Tester |
Impact pendulum test for plastics, metals and composites Energy ranging from 0.5 - 50J Capable of conducting Charpy, Izod, Pipe and Tensile Impact methods |
17 |
Fume Hood |
Lab scale approx. ( L x H x Dmm) 1120 x 680 x 1200 working volume Provision for water, inert gas, air lines and drainage system Transparent sliding front door, High efficiency fume extraction |
18 |
Digital Balance |
A balance up to five digit accuracy for DSC samples measurements, LCD display. |